Maintenance to be performed
Before every experiment
Check Sheath level (>= 1/2 bottle)
Check for fluidics / laser discrepancies in camera
alignment of sample injection ok?
alignment/look of laser beams ok?
check blue (and red if used) laser with calibration beads
very thin peaks in histograms? all ok!
broad and/or undefined peaks? Ask Till. If Till is not available, ask Michael Scholz.
Weekly
Check Waste level
Leave blue solution in device overnight (not longer!) (Warning appears on FACS log)
Routine bead measurement (Till)
Empty waste bin
Monthly
Check UV lamp runtime (<300 hours) (Warning appears on FACS log)
Check liquid level of waste canisters
3-Monthly
Check supplies (tubes, lamps, liquids)
E
dit
|
A
ttach
|
P
rint version
|
H
istory
: r3
<
r2
<
r1
|
B
acklinks
|
V
iew wiki text
|
Edit
w
iki text
|
M
ore topic actions
Topic revision: r3 - 2011-08-15,
TillDettmering
- This page was cached on 2021-01-27 - 07:51.
FACS
Log In
Toolbox
Users
Groups
Index
Search
Changes
Notifications
RSS Feed
Statistics
Preferences
User Reference
BeginnersStartHere
EditingShorthand
Macros
MacrosQuickReference
FormattedSearch
QuerySearch
DocumentGraphics
SkinBrowser
InstalledPlugins
Admin Maintenance
Reference Manual
AdminToolsCategory
InterWikis
ManagingWebs
SiteTools
DefaultPreferences
WebPreferences
Categories
Webs
Aikido
CSframework
ControlsCollaboration
CwLinac
DABC
DVApublic
EE
EPS
Epics
FACS
FAIRcontrols
FIRST
Grid
IKBestX
ITInfo
ServiceKatalog
ITRedakteure
Java
Laserionacc
Linux
Main
NIUser
PHDStudents
PLinacBD
Personalpages
ROOT
Research
SCM
SIS100BD
STF
Tasca
Subweb
Z6
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding GSI Wiki?
Send feedback
|
Imprint (german)
|
Privacy Policy (german)